Manager:
Yinghua Chen, PhD
Contact Email: yinghua.chen@case.edu
Contact Phone: 216-368-8962
Availablility: By Availability
|
Major Applications:
The multi-angle light scattering detector for determining the absolute molar masses and size of your molecules from a few hundred daltons to several million without the need for reference standards or column calibration.
Major Specifications:
Light Source: 50mW GaAs linearly polarized
Laser Wavelength: 658nm (nominal) Each instrument is individually characterized.
Laser Lifetime: >10,000 hours
Detectors: 3 angles
Detector type: Custom-designed, ultra-quiet hybrid photodiodes with built in FET transimpedance amplifiers
Detector Resolution: 24 bit
Mobile phase compatibility: All-solvent compatible (aqueous and organic)
Instrument Options: Embedded QELS at 90 degrees / Embedded COMET flow cell cleaning system
Scattering Volume: 0.02 μL
Signal Processing: On chip sinc3 filter with proprietary dust rejection filter
Electronic Inputs: 4 differential analog inputs with 24 bit resolution. TTL inputs for Alarm in, Autoinject contact closure, Recycle in
Electronic Outputs: 3 single ended analog outputs from (-10V, 10V ) which retransmit any measurement signal. TTL alarm output, autoinject retransmit, Recycle valve power
Molar Mass Range: ≤103 to 106 g/mole (daltons)*
Molecular Size Range: 10 nm to 50 nm†
Computer Interface: Ethernet and USB 2.0 interface
Processed Data Transmission Rate: Software selectable from 36.6 Hz to 0.2 Hz
Front Panel Display: 162.5mm 16-bit high resolution LCD display with Graphical User Interface
Power: 960W max, 75W typical